Copper is a very active catalyst for the low-temperature oxidation of hydrocarbons. Consequently, copper concentrations above 12 µg/kg in commercial gasoline are expected to promote the deterioration of olefins and the formation of gum.
In order to monitor heavy metal traces, anodic stripping voltammetry (ASV) is the method of choice. ASV is a sensitive and selective electroanalytical technique comprising two steps:
- metal ions are electrodeposited on the mercury electrode, which is held at a suitable potential
- the (amalgamated) metal deposits are anodically stripped from the mercury electrode by scanning the potential
The ASV is performed using the hanging mercury drop electrode (HMDE) and the Metrohm 797 VA Computrace for trace analysis. The HMDE is used with a platinum auxiliary electrode and an Ag/AgCI reference electrode using saturated lithium chloride in ethanol as the inner electrolyte. The concentration of the copper in the samples is quantified by standard addition.
|